Arginase Inhibitor Screening Service
Summary
In order to test the suitability of BioAssay Systems Arginase Assay for screening for arginase inhibitors, we performed some initial experiments to determine the Km of L-arginine for a purified human arginase and the IC50 for a known arginase inhibitor, ABH. We measured a Km of 3.4 ± 0.2 mM. Using 3 mM L-arginine, we observed an IC50 = 0.14 ± 0.1 µM for ABH. We also verified that 2 v% DMSO will not affect the assay.
Results
Km of L-Arginine
In order to determine the Km of L-arginine for Arginase I, we titrated L-arginine from 0–20 mM with 25 ng Arginase I and computed the rates of reaction, V (µmol/min/mg), for each L-arginine concentration. The rates of reaction were then plotted versus L-arginine and the Km was computed using Prism 4 (GraphPad Software Inc). We found the Km = 3.4 ± 0.2 mM.
ABH IC50 Determination
The IC50 determination required three steps: 1) 30 min pre-incubation of 25 ng arginase I with ABH, 2) arginase reaction with 3 mM L-arginine for 25 min and 3) urea concentration measurement. We titrated ABH from 0-10 µM and computed the rate of reaction, V, for each ABH concentration. The rates of reaction were then plotted versus the Log [ABH] and the IC50 was computed using Prism 4 (GraphPad Software Inc). We observed an IC50 = 0.14 ± 0.01 µM.
In order to determine the Km of L-arginine for Arginase I, we titrated L-arginine from 0–20 mM with 25 ng Arginase I and computed the rates of reaction, V (µmol/min/mg), for each L-arginine concentration. The rates of reaction were then plotted versus L-arginine and the Km was computed using Prism 4 (GraphPad Software Inc). We found the Km = 3.4 ± 0.2 mM.
ABH IC50 Determination
The IC50 determination required three steps: 1) 30 min pre-incubation of 25 ng arginase I with ABH, 2) arginase reaction with 3 mM L-arginine for 25 min and 3) urea concentration measurement. We titrated ABH from 0-10 µM and computed the rate of reaction, V, for each ABH concentration. The rates of reaction were then plotted versus the Log [ABH] and the IC50 was computed using Prism 4 (GraphPad Software Inc). We observed an IC50 = 0.14 ± 0.01 µM.
Effect of DMSO
In order to determine if DMSO would interfere with this arginase assay, we set up four reactions: 25 ng arginase I ± 2 v% DMSO and 0 ng arginase ± 2 v% DMSO. Following a 20 min reaction and addition of the Urea Reagent, we measured the OD520 and OD430 and computed the ΔOD (ODarginase – ODno arginase) for the 2 v% DMSO and 0 v% DMSO reactions at each wavelength. The following table displays the results:
Clearly 2 v% DMSO has no adverse effects on the assay and will not be an issue for the compound screen.
In order to determine if DMSO would interfere with this arginase assay, we set up four reactions: 25 ng arginase I ± 2 v% DMSO and 0 ng arginase ± 2 v% DMSO. Following a 20 min reaction and addition of the Urea Reagent, we measured the OD520 and OD430 and computed the ΔOD (ODarginase – ODno arginase) for the 2 v% DMSO and 0 v% DMSO reactions at each wavelength. The following table displays the results:
Sample | ΔOD520 | ΔOD430 |
---|---|---|
2 v% DMSO | 0.3995 | 1.4476 |
0 v% DMSO | 0.3978 | 1.4181 |
Conclusions
The results for the Km of L-arginine for arginase I and IC50 of ABH observed in this study agree well with numbers found in literature (within 3 fold). This study also verified that 2 v% DMSO will not interfere with the assay. Thus, this arginase assay will be suitable for screening for arginase inhibitors.
Please email or call us at 1-510-782-9988 x 2 to discuss your service needs.
Please email or call us at 1-510-782-9988 x 2 to discuss your service needs.
Customer Testimonials
"The guys at BioAssay Systems know their science! I had to develop a bioassay for our company and Robert was very quick to respond to keep us up to date. They are knowledgeable and very resourceful. They directed us to one of the bulk suppliers that could give us a really good deal on our active ingredient. They keep very detailed records for our research needs. I would recommend working with BioAssay Systems on your assay development projects!"
"It is a pleasure working together with BioAssay Systems. I brought samples to their Hayward location and received the results in less than 24 hr. The discussion with Robert Z was courteous, open, and with a depth of knowledge. Robert has a PhD from Stanford. Also, Frank Huang, the CEO, had sufficient scientific curiosity to try out our cleanser and provide feedback! This is beyond the call of customer relations. BioAssay Systems is a company with a lot of expertise and a friendly, responsive approach to problem solving. I will enthusiastically continue to work with them to carry out our research."
"I would like to show my appreciation to the Bioassay Systems service team. We are developing an assay to measure the change in lactate concentration of single cell samples using fluorescent microscopy. The BioAssay Systems team was quite helpful in assisting us to optimize their fluorescent L-Lactate assay kit for our unique experimental set-up. The BioAssay Systems team is always friendly and knowledgeable. We know that we can rely on them for fast and professional service."
"While developing biodegradable coatings for Drug Eluting Stents I worked with Bob to develop an accurate, precise and sensitive assay for determining the amount of polymer remaining in tissue. It was a pleasure to work with Bob and his team, they are very thorough, they understand the science very well and they always deliver on time and at the right price. I would highly recommend anyone using BioAssay Systems services in the future."
Previous
Next