Acetylcholinesterase Inhibitor Screening Service

Summary

In order to develop an Acetylcholinesterase inhibitor screening assay kit, we performed some initial experiments to establish appropriate assay conditions. We assessed conditions for Acetylcholinesterase from electric eel. First, we titrated the AChE using 1 mM Acetylthiocholine iodide as the substrate and DTNB as the chromagen. From this titration, we decided to use 0.0035 Units/reaction AChE for assay development and IC50 determinations. The Km of Acetylthiocholine iodide for AChE was then determined. We next titrated DMSO to determine the highest concentration that would be tolerated by AChE. Finally, the IC50’s for two known Acetylcholinesterase inhibitors, Donepezil and Physostigmine, were measured. We determined the Km to be 0.425 mM and the maximum DMSO concentration to be 100 v%. Using 0.425 mM acetylthiocholine iodide, we observed an IC50 = 48.3 nM for Donepezil and an IC50 = 46.7 nM for Physostigmine when each inhibitor was dissolved in 20v% DMSO.

Results

AChE Titration In order to determine an appropriate amount of AChE to use for the inhibitor screening assay and for IC50 determinations, we titrated AChE. The reaction was performed using 1 mM Acetylthiocholine Iodide and 0.4 mM DTNB dissolved in Assay Buffer pH 7.6. From these results, we opted to use 0.0035 U/rxn AChE. At this concentration, the reaction would remain linear for at least 10 minutes, and the ΔOD should be significant (>0.2) even at the Acetylthiocholine Iodide Km.

Km of Acetylthiocholine Iodide

In order to determine the Km of Acetylthiocholine iodide for Acetylcholinesterase (electric eel), we titrated acetylthiocholine iodide from 0-4.3 mM with 0.0035 U/rxn AChE and computed ΔOD (OD10min – OD0min), which is proportional to the rate of reaction, for acetylthiocholine iodide. The ΔOD’s were then plotted versus acetylthiocholine iodide concentration and the Km was computed using Prism 4 (GraphPad Software Inc.). We found the Km = 0.425 mM for Acetylthiocholine Iodide.

Acetylcholinesterase-Inhibitor-Screening-Servicefig1

DMSO Titration

In order to determine tolerance for DMSO of Acetylcholinesterase, we ran a DMSO titration (from 0-100 v% for the 5 µL of variable compound in a 200 µL reaction). For this titration we used 0.0035 U/rxn AChE. The reaction was initiated by adding Acetylthiocholine Iodide, DTNB, and Assay buffer pH 7.6 (Final concentrations 0.425 mM Acetylthiocholine Iodide and 0.3 mM DTNB) and was allowed to proceed for 10 min. We then computed the ΔOD (OD10min – OD0min), which is proportional to the rate of reaction, for each DMSO concentration. The ΔOD’s were then plotted versus DMSO concentration. From these results, we discovered that that we can use up to 100 v% DMSO if needed, but decided to move forward with 20 v% DMSO for IC50 determinations.

Acetylcholinesterase-Inhibitor-Screening-Servicefig2

Donepezil IC50 Determination

The IC50 determination required two steps: 1) 15 min pre-incubation of 45 µL AChE (78 U/L) with 5 µL Donepezil (in 20 v% DMSO), and 2) acetylcholinesterase reaction with 150 µL Working reagent containing acetylthiocholine iodide, DTNB, and assay buffer for 10 min (0.425 mM [Acetylthiocholine Iodide] and 0.31 mM [DTNB] final). We titrated Donepezil from 0-25 µM and computed the ΔOD for each Donepezil concentration. The ΔOD’s were then plotted versus the Log [Donepezil] and the IC50 was computed using Prism 4 (GraphPad Software Inc.). We observed an IC50 = 48.3 nM.

Acetylcholinesterase-Inhibitor-Screening-Servicefig3

Physostigmine IC50 Determination

The IC50 determination required two steps: 1) 15 min pre-incubation of 45 µL AChE (78 U/L) with 5 µL Physostigmine (in 20 v% DMSO), and 2) acetylcholinesterase reaction with 150 µL Working reagent containing acetylthiocholine iodide, DTNB, and assay buffer for 10 min (0.425 mM [Acetylthiocholine Iodide] and 0.31 mM [DTNB] final). We titrated Physostigmine from 0-25 µM and computed the ΔOD for each Physostigmine concentration. The ΔOD’s were then plotted versus the Log [Physostigmine] and the IC50 was computed using Prism 4 (GraphPad Software Inc.). We observed an IC50 = 46.7 nM.

Acetylcholinesterase-Inhibitor-Screening-Servicefig4
Please email or call 1-510-782-9988 x 2 to request assay service.